Portable multifunctional scintillation gamma radiation spectrometers to search, detect and identify radionuclides, measure gamma radiation energy distribution, ambient gamma radaition dose equivalent rate and alpha and beta radiation flux density
.jpg)
Features

| - Spectrometric smart probes with computer interface |
|
| - Effective search mode |
|
| - Spectrometric dose rate measuring method by transformation operator 'Spectrum-dose' |
|
| - Built-in G-M counter in the processing unit (PU) to extend the dose rate measuring range |
|
| - Dose rate and flux density measuring with automatic background subtraction |
|
| - Continuous automatic LED stabilization of the energy scale and its periodic calibration using a KCl check sample |
|
| - Digital temperature compensation of the measuring path by an internal temperature sensor |
|
| - Audible and visual alarm at searching and identifying gamma radiation radionuclides, at dose rate and flux density threshold exceeding |
|
| - Logging up to 300 spectra |
|
| - Spectrometric data readout on a backlit matrix LCD 128x64 |
|
| - Wide-temperature operation under field conditions |
|
| - Hermetic enclosure ? 121x477 mm for submerging |
|
Application

| - Environmental monitoring |
| - Radioactive waste monitoring |
| - Illicit trafficking of radiactive sources and materials monitoring |
| - Radiation monitoring of scrap metal |
| - Nuclear industry |
| - Geological survey |
| - Nuclear medicine |
| - Scientific research |
| - Emergency |
| IDENTIFICATION |
| Natural (NORM) - 40K, 232Th, 238U |
| Industrial - 241Am, 133Ba, 57Co, 60Co, 137Cs, 192Ir, 226Ra, 228Th, 22Na, 54Mn, 152Eu, 75Se |
| Medical - 67Ga, 123I, 125I, 131I, 111In, 99mTc, 201Tl, 133Xe, 51Cr |
Specification
| |
 |
 |
| Detectors |
|
| AT6101 (BDKG-05) |
NaI (Tl), ? 40х40 mm |
| AT6101B (BDKG-11) |
NaI (Tl), ? 63х63 mm |
| PU |
G-M counter CI-29BG |
| Gamma radiation detection in energy ranges |
|
| BDKG-05, BDKG-11 |
20 - 1500 and 40-3000 keV |
| Alpha radiation detection in energy range |
|
| BDPA-01 |
4-7 MeV |
| Beta radiation detection with maximum energies |
|
| BDPB-01 |
from 155 keV (14C) to 3.5 MeV (106Ru+106Rh) |
| Integral non-linearity |
not more than ± 1 % |
| Relative energy resolution on 137Cs |
|
| BDKG-05 |
not more than 9 % |
| BDKG-11 |
not more than 9.5 % |
| Maximum input statistic load |
not less than 5·104 s-1 |
| Continuous operation time with built-in accumulator unit |
not less than 12 h |
| Energy scale instability within continuous operation of 12 h |
not more than 1 % |
| Dose rate measuring range |
|
| BDKG-05 |
0.01 - 300 µSv/h |
| BDKG-11 |
0.01 - 100 µSv/h |
| PU |
1 µSv/h - 10 mSv/h |
| Energy sensitivity response respect to 137Cs: |
|
| BDKG-05, BDKG-11 (50-3000 keV) |
± 20 % |
| PU (60-3000 keV) |
± 35 % |
| Flux density measuring range |
|
| BDPA-01: alpha radiation from surfaces |
0,5 - 105 part./(min·cm2) |
| BDPB-01: beta radiation from surfaces |
3 - 5·105 part./(min·cm2) |
| Intrinsic measurement error |
|
| dose rate |
not more than ± 20 % |
| flux density |
not more than ± 20 % |
| Spectrum acquiring time |
from 1 to 64800 s |
| Sensitivity |
|
| 241Am |
|
| BDKG-05 |
5600 cps/µSv·h-1 |
| BDKG-11 |
12700 cps/µSv·h-1 |
| 137Cs |
|
| BDKG-05 |
670 cps/µSv·h-1 |
| BDKG-11 |
1960 cps/µSv·h-1 |
| 60Co |
|
| BDKG-05 |
330 cps/µSv·h-1 |
| BDKG-11 |
1030 cps/µSv·h-1 |
| radiation background 0.08 µSv/h |
|
| BDKG-05 |
100 cps |
| BDKG-11 |
270 cps |
| Operating temperature range |
-20°С - +50°С |
| Relative air humidity at 35oC |
95 % |
| Operation mode setup time |
not more than 10 min |
| Protection class |
IP54 |
| Radio disturbance |
|
| CEI/IEC CISPR 22:1997 |
|
| Electromagnetic compatibility |
|
| CEI/IEC 61000-4-2:1995 |
|
| IEC 61000-4-3:1995 |
|
| Weight |
|
| BDKG-05 |
1.2 kg |
| BDKG-11 |
1.9 kg |
| BDPA-01 |
0.5 kg |
| BDPB-01 |
0.5 kg |
| PU |
0.8 kg |
| Dimensions |
|
| BDKG-05 |
? 60х320 mm |
| BDKG-11 |
? 80х345 mm |
| BDPA-01 |
? 80х196 mm |
| BDPB-01 |
? 80х196 mm |
| PU |
?109х220х35 mm |

Complete set: spectrometric gamma radiation smart probe, processing unit, AC adapter, check sample, shoulder strap, Manual, paking case and package.
External alpha and beta radiation smart probes, telescopic bar 1.6 m, hermetic enclosure, kit of accessories to connect to PC and applied software to acquire and process spectra on PC are options and they are supplied on additional order.
The spectrometers AT6101 and AT6101B comply with IEC 62327 International standard requirements. They also conform with the 89/336/EEC Directive complying with EN 61000-6-3 and EN 61000-6-2 standard requirements.
|